Product Details:
Model Name/Number | A - 396A |
Brand | Dwyer |
Weight | 14 oz (400 g). |
Working Pressure | 72 psi (5 bar) |
Temperature | -20 to 120 Degree F (-29 to 49 Degree C) |
Output Connection | 1/8 " NPT(F |
Dwyer Calibration Pump Model A-396A
Model A-396A Calibration Pump Generates up to 72 psig, Integral Bleed Fitting Serves as Pressure Source to calibrate Gages and Transmitters or to Set Pressure Switches. Use with Manometer or other Pressure Standard. Includes volume adjuster enabling fine pressure control and bleed valve. Generates pressures from a fraction of an inch w.c. to 72 psig (5 bar). Includes barbed fitting, tee connector and three 36" lengths of vinyl tubing.
Model - A-396ADescription - Calibration Pump, serves as pressure source to calibrate gages and transmitters or to set pressure switches. Use with manometer or other pressure standard. Includes volume adjuster enabling fine pressure control and bleed valve.Specifications - Installation and Operating InstructionsPHYSICAL DATA :Working Pressure: 72 psi (5 bar).Temperature Limits: -20 to 120°F (-29 to 49°C).Displaced Volume: Approximately 1.2 in3(20 cm3) with 3/64 ?? (1 mm)thread revolution (turn).Output Connection: 1/8 ?? NPT(F).Overall Dimensions:7-7/8 ?? (20 cm) length, 1-3/8 ?? (3.5 cm) diameter.Weight:14 oz (400 g).INSTALLATION
CAUTION: Do Not Over Tighten The Fitting. Damage to the threads in thecylinder or interface with Moving Parts inside the Cylinder may Result.
Connect the calibration pump to both the manometer (or other pressurestandard of your choice) and the device to be calibrated, by means of atee connector as shown in Figure A. Low pressure tubing is provided. Tomaintain pressure integrity, the use of Teflon® tape to seal the pipethreads is recommended. Referring to Figure B, close the vent valve (1).Holding the pump body in one hand and grasping the pump tee handle(2) in the other hand, gently pump until the desired pressure is obtained.Fine adjustment of the pressure can be achieved by turning the vernierknob (3) clockwise to increase pressure and counter-clockwise Todecrease Pressure. If the desired calibration pressure is less than a fewinches of water column, use only the vernier knob (3) to achieve thedesired calibration pressure. After the Calibration test is complete, openvent valve (1) to Release the pressure.
NOTE:
For Pressures Above 40 psi, use a 1/8 ?? NPT (M) Compressionfitting and appropriately sized rigid tubing (not included) in Place Ofbarbed Fitting and Flexible Tubing
PHYSICAL DATA:
Working Pressure: 72 psi (5 bar).
Temperature Limits: -20 to 120°F (-29 to 49°C).
Displaced Volume:
Approximately 1.2 in3(20 cm3) with 3/64 ?? (1 mm)thread revolution (turn).
Output Connection: 1/8 ?? NPT(F).
Overall Dimensions:7-7/8 ?? (20 cm)
length, 1-3/8 ?? (3.5 cm) diameter.
Weight:14 oz (400 g).
PHYSICAL DATA:
Working Pressure: 72 psi (5 bar).
Temperature Limits: -20 to 120°F (-29 to 49°C).
Displaced Volume: Approximately 1.2 in3(20 cm3) with 3/64 ?? (1 mm)thread revolution (turn).Output Connection: 1/8 ?? NPT(F).Overall Dimensions:7-7/8 ?? (20 cm) length, 1-3/8 ?? (3.5 cm) diameter.Weight:14 oz (400 g).
Additional Information:
Product Details:
Model Name/Number | Model A-396 A |
Brand | Dwyer |
Working Pressure | 72 psi (5 bar) |
Temperature | -20 to 120 Degree F (-29 to 49 Degree C) |
Output Connection | 1/8 " NPT(F |
Length | 7-7/8 " (20 cm) |
Diameter | 1-3/8 " (3.5 cm) |
Weight | 14 oz (400 g). |
Product Details:
Minimum Order Quantity | 1 Piece |
Brand | Dwyer |
Weight | 14 oz (400 g). |
Working Pressure | 72 psi (5 bar) |
Temperature | -20 to 120 Degree F (-29 to 49 Degree C) |
Output Connection | 1/8 " NPT(F) |
Length | 7-7/8 " (20 cm) |
Diameter | 1-3/8 " (3.5 cm) |
Application:
Additional Information: